Publication:
Numerical and experimental study of the fluid flow through a medical device

Date

2015

Authors

Nicolás, M.
Palero, V. R.
Peña, Estefanía
Arroyo, M. P.

Director

Publisher

Elsevier
Acceso abierto / Sarbide irekia
Artículo / Artikulua
Versión aceptada / Onetsi den bertsioa

Project identifier

Abstract

The purpose of this paper is to verify a commercial software based fluid–structure interaction scheme for the inferior vena cava. Vena cava deep thrombosis (TVP) is a potentially deathly disease consequent to pulmonary thromboembolism (TEP). TEP consist in the obstruction of the pulmonary artery due to a blood clot traveling in the cardiovascular system and is treated with anticoagulants and inferior vena cava filters. Flow fields along the vena cava and an antithrombus filter were studied and compared with a Particle Image Velocimetry (PIV) based model to validate the numerical model. The results show that the fluid–structure interaction (FSI) models are valid and can be used to study the deformations in the inferior vena cava wall using patient-specific geometries.

Description

Keywords

Inferior vena cava, FSI, CFD, PIV, Antithrombus filter

Department

Ingeniería Mecánica, Energética y de Materiales / Mekanika, Energetika eta Materialen Ingeniaritza

Faculty/School

Degree

Doctorate program

item.page.cita

item.page.rights

© 2014 Elsevier Ltd. This manuscript version is made available under the CC-BY-NC-ND 4.0.