Publication: Luminance calibration of a full sky HDR imaging system using sky scanner measurements
Date
Director
Publisher
Abstract
A full sky High Dynamic Range imaging system, based on a Single-Lens Reflex camera with a fisheye lens, has been constructed and calibrated with a sky scanner luminance meter. The method considers the geometrical, spectral, timing and orientation issues between instruments. The calibration data sets, having nearly simultaneous measurements under stable sky conditions, were obtained from approximately one month of data using selection variables based in the experimental design. For luminance estimation we use the standard 𝐶𝐼𝐸𝑌 RGB combination and a Spectrally Matched Luminance (𝑆𝑀𝐿) predictor, matching the spectral response of the instruments. With 738 calibration points having luminances up to 23.6 kcd∕m2, covering 98.5% of the sky luminance range, 𝐶𝐼𝐸𝑌 is linearly correlated with sky scanner measurements with a coefficient of determination 𝑅2 = 0.9927 and a Root Mean Squared Error (𝑅𝑀𝑆𝐸) of 7.7%. 𝑆𝑀𝐿 gives better results, with 𝑅2 = 0.9973 and 𝑅𝑀𝑆𝐸 = 5.3%. With 253 calibration points with luminances up to 12.9 kcd∕m2, comprising 94.1% of the sky luminance range, both predictors clearly improve, with 𝑅2 = 0.9964 and 𝑅𝑀𝑆𝐸 = 4.1% in case of 𝐶𝐼𝐸𝑌 and 𝑅2 = 0.9982 and 𝑅𝑀𝑆𝐸 = 2.9% in case of 𝑆𝑀𝐿.