In vitro comparison of Günther Tulip and Celect filters. Testing filtering efficiency and pressure drop
dc.creator | Nicolás, M. | es_ES |
dc.creator | Malvè, Mauro | es_ES |
dc.creator | Peña, Estefanía | es_ES |
dc.creator | Martínez, Miguel Ángel | es_ES |
dc.creator | Leask, R. | es_ES |
dc.date.accessioned | 2020-09-16T11:21:46Z | |
dc.date.available | 2020-09-16T11:21:46Z | |
dc.date.issued | 2015 | |
dc.identifier.issn | 0021-9290 | |
dc.identifier.uri | https://hdl.handle.net/2454/38106 | |
dc.description.abstract | In this study, the trapping ability of the Günther Tulip and Celect inferior vena cava filters was evaluated. Thrombus capture rates of the filters were tested in vitro in horizontal position with thrombus diameters of 3 and 6 mm and tube diameter of 19 mm. The filters were tested in centered and tilted positions. Sets of 30 clots were injected into the model and the same process was repeated 20 times for each different condition simulated. Pressure drop experienced along the system was also measured and the percentage of clots captured was recorded. The Günther Tulip filter showed superiority in all cases, trapping almost 100% of 6 mm clots both in an eccentric and tilted position and trapping 81.7% of the 3 mm clots in a centered position and 69.3% in a maximum tilted position. The efficiency of all filters tested decreased as the size of the embolus decreased and as the filter was tilted. The injection of 6 clots raised the pressure drop to 4.1 mmHg, which is a reasonable value that does not cause the obstruction of blood flow through the system. | en |
dc.description.sponsorship | The authors gratefully acknowledge research support from the Spanish Ministry of Science and Technology through the research project DPI2010-20746-C03-01, and the CIBER initiative. We also thank the Spanish Ministry for the financial support to M. Nicolas through the grant BES-2011-043471 and the grant EEBB-I-13-07507 for the research stage in Montreal. | en |
dc.format.extent | 53 p. | |
dc.format.mimetype | application/pdf | en |
dc.language.iso | eng | en |
dc.publisher | Elsevier | en |
dc.relation.ispartof | Journal of Biomechanics, 2015, 48(3), 504-511 | en |
dc.rights | © 2014 Elsevier Ltd. This manuscript version is made available under the CC-BY-NC-ND 4.0. | en |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | |
dc.subject | Pulmonary embolism | en |
dc.subject | Inferior vena cava | en |
dc.subject | Günther Tulip filter | en |
dc.subject | Celect filter | en |
dc.subject | Clot capture rate | en |
dc.subject | Filter efficiency | en |
dc.title | In vitro comparison of Günther Tulip and Celect filters. Testing filtering efficiency and pressure drop | en |
dc.type | info:eu-repo/semantics/article | en |
dc.type | Artículo / Artikulua | es |
dc.contributor.department | Ingeniería Mecánica, Energética y de Materiales | es_ES |
dc.contributor.department | Mekanika, Energetika eta Materialen Ingeniaritza | eu |
dc.rights.accessRights | info:eu-repo/semantics/openAccess | en |
dc.rights.accessRights | Acceso abierto / Sarbide irekia | es |
dc.identifier.doi | 10.1016/j.jbiomech.2014.11.047 | |
dc.relation.publisherversion | https://doi.org/10.1016/j.jbiomech.2014.11.047 | |
dc.type.version | info:eu-repo/semantics/acceptedVersion | en |
dc.type.version | Versión aceptada / Onetsi den bertsioa | es |