Publication: Strategies for fabrication of hydrogen peroxide sensors based on electrostatic self-assembly (ESA) method
Consultable a partir de
Date
Director
Publisher
Project identifier
Abstract
An optical fiber sensor has been fabricated for detection of 5 M–0.1mM concentrations of hydrogen peroxide with a linear response. The deposition method used is electrostatic self-assembly (ESA) of polymer cationic and anionic layers. Prussian Blue (PB) has been included in the polycation layers. The optical fiber sensor is included in a reflection setup, where the measuring technique is based on the slope of the optical reflected power change caused by oxidation of Prussian White (PW) to Prussian Blue. The sensor recovers after immersion in a reductive agent and is immune against a variety of components. Measurement of hydrogen peroxide has been proved successfully in a wide range pHs between 3 and 9. Some techniques have been applied in order to avoid the lost of indicator.
Keywords
Department
Faculty/School
Degree
Doctorate program
Editor version
Funding entities
© 2004 Elsevier B.V. The manuscript version is made available under the CC BY-NC-ND 4.0 license